BT100-1L Peristaltic Pump Remote Control Function
Release Time:2016-12-09
Communication Control
The pump can connect to control computer (computer, PLC, SCM)through RS485 serial communication interface (shown as below).
Analog Control
Set the “Remote Control” in “on” state. External Control Module is shown as below.
Terminal Definition of Standard External Control Module
1# Analog speed control input, for external speed control in external control modules of 4 - 20 mA, 0 - 5V and 0 - 10 V.
2# External start/stop control input. When connected to COM,the pump runs, when connected to high level (5 -12 V), the pump stops.
3#External cw/ccw control input. When connected to COM, the pump rotates cw, when connected to high level (5 -12 V), the pump rotates ccw.
4# Reference potential for 4 - 20 mA, 0 - 5V/10V speed control
input.
5# Reference potential for external cw/ccw, external start/stop input.