Home> Technical Support > Syringe Pump > Syringe Pump Video

MSP1 CX Valve & Syringe Installation


Technical Specifications

Main Spec
Accuracy:≤1.0 % (syringe ≥1 mL)
Precision:CV≤0.05% at full stroke (with syringe ≥ 500uL)
Wetted material:Borosilicate glass,PTFE,PFA,CTFE,ETFE,UHMW-PE
Dimension (L x W x H):114mm x 65mm x 254mm
Weight:2.15 kg
Power supply:DC 24V,1.5A (peak)

Syringe and Drive
Syringe fitting:1/4″-28 thread
Syringe:50uL, 100uL, 250uL, 500uL, 1mL, 2.5mL, 5mL, 10mL, 25mL
Syringe material:Borosilicate glass,stainless steel plunger with PTFE coating, PTFE or UHMW-PE plunger tip
Resolution (standard mode):6000 steps with 0.01mm in 1 step
Resolution (microstep mode):48000 microsteps with 1.25um in 1 microstep
Travel:60mm
Time for full stroke:1.26s-1200s
Drive system:Ball screw drive with optical encoder

Valve & Valve Drive
Turn Time:≤250ms between adjacent ports
Valve options:Standard: 3-port valve, optional: 3-port distribution valve,  4-port valve, T-valve, Y-block
Max pressure:0.68MPA
Fitting:1/4″-28 thread

Communication
Interface:RS485, RS232, CAN bus,external control input and output
Protocol:Data terminal,OEM protocol (serial)  or  CAN
Baud rate:"RS485 or RS232: 9600bps or 38400bps
CAN bus: 100 KB, 125 KB 250 KB, 500 KB or 1MB"
Format:Data bit: 8;  parity: none; stop bit: 1;  half duplex (RS232/485)

Firmware
Topology group:Up to 15 pumps in one group, support group control and individually addressed
Upgrade firmware online:"One Click"  by customer  through PC Utility
Homing Algorithm:Plunger home position is identified by motor stall, which will reduce the dead volume of the system
Halt function:For better interactive control, "h " command is used to halt execution of the command string during plunger movement. To resume execution with "R" or "r" command
Microstep mode:Set microstep mode for smoother motion with 48000 microsteps per stroke, 1/8 microstepping or 1/16 microstepping
Programmable parameter:Programmable ramps, cut-off velocity, plunger speeds, backlash compensation,  programmable work-flow command set , delays and loops, terminate moves, diagnostics, absolute and relative positions

Environmental
Operating temperature:Recommended: 15℃-40℃ ,use low temperature syringe for low temperature application
Operating Humidity:<80% RH,  no condensation
Storage temperature: -20℃-65℃
Storage Humidity:<80% RH, no condensation


<< Back