BT300-1F Peristaltic Pump Filling and Dispensing Setting
Release Time:2016-12-16
Enter “Fill Setting” interface:
1. Other parameters can’t be set here. Set the other parameters first and then go along the “Fill Setting”. Other parameters setting methods please see page 8 “System Setting”.
2. In “Running Interface” press the Adjusting Knob to enter “Fill Setting” interface. Press the Adjusting Knob again to enter filling setting menu. Turn the Adjusting Knob to select parameters. There are 4 parameters. They are Volume, Pause time, Copy No., Flow.
Volume: In “Fill Setting” interface, press the Adjusting Knob to enter the next interface. Press the Adjusting Knob to enter “Volume” interface. Turn the Adjusting Knob to input required volume. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the setting and return to the previous interface.
Pause Time: The pause time between two filling. Its range is 1s – 999s. In “Fill Setting” interface, press the Adjusting Knob to enter the next interface. Turn the Adjusting Knob to highlight the “Pause time”. Press the Adjusting Knob to enter “Pause time” interface. Turn the Adjusting Knob to adjust the pause time. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the setting and return to the previous interface.
Copy No.: the total filling number. Its range is 1 – 999.
In “Fill Setting” interface, press the Adjusting Knob to enter the next interface. Turn the Adjusting Knob to highlight the “Copy No.”. Press the Adjusting Knob to enter “Copy No.” interface. Turn the Adjusting Knob to adjust the copy number. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the setting and return to the previous interface.
Flow: Adjust the dispensing and filling efficiency. Its range is determined by the pump head and tubing.
In “Fill Setting” interface, press the Adjusting Knob to enter the next interface. Turn the Adjusting Knob to highlight the “Flow”. Press the Adjusting Knob to enter “Flow” interface. Turn the Adjusting Knob to adjust the flow rate and dispensing time. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the setting and return to the previous interface.
Disp: Filling time for one time.
Note:
Change the tubing if can’t get the desired Disp. time.
Calibration Function
Calibration is needed in below situation:
1. Use the pump the first time; 2. Change the pump; 3. Change the tubing; 4. Setting new flow rate; 5. Operating condition changes.
Enter calibration function setting interface:
In “Running Interface”, press the Adjusting Knob to enter “Fill Setting” interface, turn the Adjusting Knob to enter “Calibration” interface.
Flow Rate Calibration
In “Calibration” interface, press the Adjusting Knob to enter “Test Time” setting interface. Turn the Adjusting Knob to set the test time. Its range is 1s – 999s. Press the Adjusting Knob to enter the next interface. Press the Start/Stop Key to begin the testing. Turn the Adjusting Knob to input the actual volume after testing. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the calibration and return to the “Calibration” interface.
Prompt:
In “Disp” state, the testing time of flow rate calibration can be set the time of disp. Time (please see the “Flow” setting of “Filling setting”) for more accurate filling volume.
1. Other parameters can’t be set here. Set the other parameters first and then go along the “Fill Setting”. Other parameters setting methods please see page 8 “System Setting”.
2. In “Running Interface” press the Adjusting Knob to enter “Fill Setting” interface. Press the Adjusting Knob again to enter filling setting menu. Turn the Adjusting Knob to select parameters. There are 4 parameters. They are Volume, Pause time, Copy No., Flow.
Volume: In “Fill Setting” interface, press the Adjusting Knob to enter the next interface. Press the Adjusting Knob to enter “Volume” interface. Turn the Adjusting Knob to input required volume. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the setting and return to the previous interface.
Pause Time: The pause time between two filling. Its range is 1s – 999s. In “Fill Setting” interface, press the Adjusting Knob to enter the next interface. Turn the Adjusting Knob to highlight the “Pause time”. Press the Adjusting Knob to enter “Pause time” interface. Turn the Adjusting Knob to adjust the pause time. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the setting and return to the previous interface.
Copy No.: the total filling number. Its range is 1 – 999.
In “Fill Setting” interface, press the Adjusting Knob to enter the next interface. Turn the Adjusting Knob to highlight the “Copy No.”. Press the Adjusting Knob to enter “Copy No.” interface. Turn the Adjusting Knob to adjust the copy number. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the setting and return to the previous interface.
Flow: Adjust the dispensing and filling efficiency. Its range is determined by the pump head and tubing.
In “Fill Setting” interface, press the Adjusting Knob to enter the next interface. Turn the Adjusting Knob to highlight the “Flow”. Press the Adjusting Knob to enter “Flow” interface. Turn the Adjusting Knob to adjust the flow rate and dispensing time. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the setting and return to the previous interface.
Disp: Filling time for one time.
Note:
Change the tubing if can’t get the desired Disp. time.
Calibration Function
Calibration is needed in below situation:
1. Use the pump the first time; 2. Change the pump; 3. Change the tubing; 4. Setting new flow rate; 5. Operating condition changes.
Enter calibration function setting interface:
In “Running Interface”, press the Adjusting Knob to enter “Fill Setting” interface, turn the Adjusting Knob to enter “Calibration” interface.
Flow Rate Calibration
In “Calibration” interface, press the Adjusting Knob to enter “Test Time” setting interface. Turn the Adjusting Knob to set the test time. Its range is 1s – 999s. Press the Adjusting Knob to enter the next interface. Press the Start/Stop Key to begin the testing. Turn the Adjusting Knob to input the actual volume after testing. Press the Adjusting Knob to confirm, save and quit. Press the Exit Key to cancel the calibration and return to the “Calibration” interface.
Prompt:
In “Disp” state, the testing time of flow rate calibration can be set the time of disp. Time (please see the “Flow” setting of “Filling setting”) for more accurate filling volume.