Programmable Syringe Pump TJP-3A Flow Rates Calibration

Release Time:2017-01-20

Flow calibration: Operator can get more accurate flow through calibrating
the diameter of syringe barrel.
In work interface and stop state, press RCS to enter menu setting interface,
turn RCS to highlight “Calibration”, press RCS to enter next interface, the
first line display the programmed flow in current mode; the second line
displays the testing time whose range is from 1 to 60 minutes. The third line
displays theoretical volume (the product of flow and testing time) or input
measured volume directly. The fourth line displays the start/stop and
direction state.
Calibration procedure: set testing time first, start pump. Pump automatically
stop after the testing time is over, measure the liquid volume and input the
measured volume, the calibration is over. If knew the measured volume in
advance, operator can directly input the measured volume.
Setting example is shown as below: