BT100-1F Peristaltic Pump Calibration
Release Time:2016-12-16
Pump should be calibrated when flow rate or dispense volume is beyond the tolerance. Flow rate and dispensing volume can be calibrated separately in different application.
*Flow Rate Calibration
1. Press and turn Knob to highlight calibration line. Press Knob to enter next interface. Turn Knob to select Flowrate. Press Knob to highlight test time line. Press Knob again to highlight the value of time. Turn Knob to select test time (its range is from 0.5 to 30 minutes). Press Knob to confirm or press Exit Key to cancel the selection and return to previous menu.
2. Press Start/Stop Key, the LCD displays running time and actual volume. Running time counts down.
3. Turn Knob to input actual tested volume. Press Knob to confirm. The calibration operation can be repeated.
If the actual volume is known in advance, enter the calibration interface and input the actual volume directly.
*Flow Rate Calibration
1. Press and turn Knob to highlight calibration line. Press Knob to enter next interface. Turn Knob to select Flowrate. Press Knob to highlight test time line. Press Knob again to highlight the value of time. Turn Knob to select test time (its range is from 0.5 to 30 minutes). Press Knob to confirm or press Exit Key to cancel the selection and return to previous menu.
2. Press Start/Stop Key, the LCD displays running time and actual volume. Running time counts down.
3. Turn Knob to input actual tested volume. Press Knob to confirm. The calibration operation can be repeated.
If the actual volume is known in advance, enter the calibration interface and input the actual volume directly.