WT600-4F Flow Mode Operation
Release Time:2016-12-30
*Enter Flowrate Mode
Follow below diagram, enter ‘Control Mode’ interface, press ‘Increment Key’ or ‘Decrement Key’ to highlight ‘Flowrate’ , press ‘Confirm Key’ to save the setting or press ‘Return Key’ to cancel the setting and back to previous menu.
*Flowrate Setting
In ‘Flowrate’ mode, press ‘Increment Key’ or Decrement Key’ to change the flow rates of the pumps.
*Flowrate Calibration:
1.Calibration Time Setting
Follow the below diagram, enter ‘Calibration’ interface under flowrate mode, press ‘Increment Key’ or ‘Decrement Key’ to adjust testing time, the range of testing time is from 30 seconds to 1800 seconds, press ‘Confirm Key’ to save the setting or press ‘Return Key’ to cancel the setting and back to previous menu.
2.Calibration Procedure
Press ‘Start/Stop Key to start flow calibration, the screen will display ‘Dispensing Time’ and ‘Actual Volume’; ‘Dispensing Time’ will count down until the procedure finish. Press ‘Increment Key’ or ‘Decrement Key’ to input actual volume, press ‘Confirm Key’ to save the setting. This procedure can be repeated in order to get a more accurate result.
3.Fast Calibration:
If the actual volume is known in advance, enter the calibration interface input the actual volume directly.
Follow below diagram, enter ‘Control Mode’ interface, press ‘Increment Key’ or ‘Decrement Key’ to highlight ‘Flowrate’ , press ‘Confirm Key’ to save the setting or press ‘Return Key’ to cancel the setting and back to previous menu.
*Flowrate Setting
In ‘Flowrate’ mode, press ‘Increment Key’ or Decrement Key’ to change the flow rates of the pumps.
*Flowrate Calibration:
1.Calibration Time Setting
Follow the below diagram, enter ‘Calibration’ interface under flowrate mode, press ‘Increment Key’ or ‘Decrement Key’ to adjust testing time, the range of testing time is from 30 seconds to 1800 seconds, press ‘Confirm Key’ to save the setting or press ‘Return Key’ to cancel the setting and back to previous menu.
2.Calibration Procedure
Press ‘Start/Stop Key to start flow calibration, the screen will display ‘Dispensing Time’ and ‘Actual Volume’; ‘Dispensing Time’ will count down until the procedure finish. Press ‘Increment Key’ or ‘Decrement Key’ to input actual volume, press ‘Confirm Key’ to save the setting. This procedure can be repeated in order to get a more accurate result.
3.Fast Calibration:
If the actual volume is known in advance, enter the calibration interface input the actual volume directly.