Home> Technical Support > Peristaltic Pump > Technical Articles

How to use External Control Function for BQ50-1J

Release Time:2016-11-28

Analog Control

Externalcontrol module is shown as below. One side is DB15 connecting peristaltic pump.Another side is wiring terminal connecting control signal. When using externalcontrol function just insert external control module to the external control interfacewhich is in the rear panel of peristaltic pump and connect the correspondingcontrol signal wire.

 Terminal Definitionof Standard External Control Module

1#N/A

2#External start/stop control input. When connectedto COM,the pump runs; When connected to high level (5 - 12V), the pumpstops.

3#External cw/ccw control input. When connected toCOM, thepump rotates cw; When connected to high level (5- 12V), thepump stops.

4#Pulse speed control input, for external speedcontrol inexternal control module of 0 - 10 KHz.

5#Reference potential for external start/stop,cw/ccw and pulsespeed control input.