LSP02(10)-1B Dual Channels Syringe Pump Flowrate setting

Release Time:2017-01-06

In parameters setting interface, turn and press RES to highlight the infusion flowrate or withdrawal flowrate setting line. Press RES to make the numerical value position highlighted. Enter the flowrate required from the membrane keypad. Press RES to set flowrate unit. Turn RES to scroll through the list and press RES to select the unit required. Possible flowrate units are µl /h, µl/m, ml/h, ml/m. The highlighted display indicates that this parameter can be changed. 


Note: If the number entered exceeds the maximum flowrate possible then the pump prompts the maximum feasible flowrate at the Time position. To continue enter a flowrate smaller than the maximum.