Email:
 
Unit / Name:
 
Home> Products > Product by Series > Syringe Pump > Laboratorial Syringe Pump

Download the data needs of Member permissions,Please Login

New to Longer? Sign Up

LSP10-1B - 10 Channels Syringe Pump

LSP10-1B - 10 Channels Syringe Pump

LSP10-1B is an infusion/withdrawal syringe pump, which could be mounted with 10 syringes at most. The acceptable syringe specification is from 10 µL to 10 mL. Suitable for high accuracy and small flow rate liquid transferring.

Flow rates:0.831nl/min-21.675ml/min

Max. Number of Channels:10


Applications:

For high precision fluid dosing, especially for laboratorial application.



Product Instruction:

LSP10-1B is an infusion/withdrawal syringe pump, which could be mounted with 10 syringes at most. The acceptable syringe specification is from 10 µL to 10 mL. Suitable for high accuracy and small flow rate liquid transferring.
CE certified product based on request.

Main Functions and Features

- Syringe Selection: The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly
- Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt.
- Working Mode: Infusion, withdrawal, infusion/withdrawal, withdrawal/infusion, continuous
- Memory Function: 1. The parameters are saved in EEPROM. The parameters don't need to be reset when power returns after an interruption. 2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns
- Protection Function: The pump will stall and give an alarm when the drive structure of the pump is blocked
- Communication Function: Realize computer control through RS485 communication interface
- External Control Function: Input/output control
- Calibration Function: Acquire accurate volume through calibration
- Syringe Protection: Adjust syringe rest to prevent syringe from damaging

Technical Specifications

Syringe Pump LSP10-1B
Number of Channels 10
Injected Volume per Micro-step 0.0257µL (10 mL BD Syringe)
Syringe Size 10µL – 10 mL
Flow Rate (60 mL Syringe) 0.831nL/min – 21.675 mL/min
Dimensions (mm) 280×250×140
Weights (Kg) 4.3
Working Mode Infusion / Withdraw
Linear Force > 18Kg
Advance per Micro-step 1/16 Step (0.156µL)
Max. Step Rate 13867 (1/16 step) /sec
Min. Step Rate 16 (1/16) / 30 sec
Max. Linear Rate 130 mm/min
Min. Linear Rate 5µL/min
Accuracy ≤ 0.5% error in the condition of ≥ 30% of max. infusion distance
Setting Mode Rotary coded switch and membrane keypad
Display 128x64 graphic LCD
Power Supply AC100 -240V
Operation Condition Temperature: 5℃ - 40℃, Relative humidity: <80%
Dimensions 280×330×140 (mm)
Weight 5.3kg

Other parameters

Type Product Code Syringe Syringe ID (mm) Flow Rate (μl/min-ml/min)
LSP10-1B 0503461 10μl 0.50 0.001-0.0255
25μl 0.80 0.0025-0.0653
50μl 1.10 0.0048-0.1235
100μl 1.60 0.0101-0.2614
250μl 2.30 0.0208-0.5401
500μl 3.25 0.0415-1.0784
1ml 4.72 0.0875-2.2747
2ml 9.00 0.3181-8.2702
5ml 13.10 0.6739-17.522
10ml 16.60 1.0821-28.135

LSP02(10)-1B Syringe Pump External control port

Pin definition 3 COM, ground ref. 8 Trigger, TTL input, falling edge start/stop pump, e.g. footswitch 4 Gate, TTL input, change from high to low - starts; when running stays low, change to high - s...

2017-01-06

LSP02(10)-1B Syringe Pump RS485 setup

The RS485 connections are made through terminal connectors, located on the rear panel. A single PC can control up to 30 pumps. When controlled by a PC each pump must be assigned an individual address....

2017-01-06

LSP02(10)-1B Syringe Pump External control function

External control function, including run/stop control input and direction control input can be enabled or disabled.1.In parameters setting interface, press and turn RES to highlight the external contr...

2017-01-06

LSP02(10)-1B Syringe Pump Calibration

The flowrate can be calibrated for more accurate dispense volume.1.In parameters setting interface, press and turn RES to highlight the Calibration line. Press RES to enter calibration interface.

2017-01-06

LSP02(10)-1B Dual Channels Syringe Pump Pause time setting

In infusion/withdrawal mode, the pause time is set in infusion parameters setting interface. In withdrawal mode, the pause time is set in withdrawal parameters setting interface. In continuous mode, t...

2017-01-06

LSP02(10)-1B Dual Channels Syringe Pump Flowrate setting

In parameters setting interface, turn and press RES to highlight the infusion flowrate or withdrawal flowrate setting line. Press RES to make the numerical value position highlighted. Enter the flowra...

2017-01-06

LSP02(10)-1B Dual Channels Syringe Pump Volume setting

In parameters setting interface, turn and press RES to highlight the infusion volume or withdrawal volume setting line. Press RES to make the numerical value position highlighted. Enter targeted volum...

2017-01-06

Syringe pump selection guides

A.How to select syringe? Choosing the proper specification according to the volume of dispensing fluid. B.How to select Dirve? Application dispensing mode of the fluid. Application environment and ins...

2015-12-19

Laboratory Syringe Pump in Laboratories

A laboratory is a facility that provides controlled conditions for scientific or technological research, where experiments and measurements can be performed. Laboratories can be divided into 3 main ...

2015-09-11

  • LSP02-2B LSP10-1B Brochure Download

  • laboratorial syringe pump LVD CE Certificate Download

  • laboratorial syringe pump ECD CE Certificate Download