Home> Technical Support > Syringe Pump > Technical Articles

Technical Articles

  • Programmable Syringe Pump TJP-3A Flow Rates Calibration

    01/20/2017

    Flow calibration: Operator can get more accurate flow through calibrating the diameter of syringe barrel. In work interface and stop state, press RCS to enter menu setting interface, turn RCS to highl...

  • Programmable Syringe Pump TJP-3A External Control

    01/20/2017

    Control the Start/Stop of the pump by external signal, the trigger and gate modes are optional. In work interface and stop state, press RCS to enter menu setting interface turn RCS to highlight “EXT ...

  • LSP02(10)-1B Syringe Pump External control port

    01/06/2017

    Pin definition 3 COM, ground ref. 8 Trigger, TTL input, falling edge start/stop pump, e.g. footswitch 4 Gate, TTL input, change from high to low - starts; when running stays low, change to high - s...

  • LSP02(10)-1B Syringe Pump RS485 setup

    01/06/2017

    The RS485 connections are made through terminal connectors, located on the rear panel. A single PC can control up to 30 pumps. When controlled by a PC each pump must be assigned an individual address....

  • LSP02(10)-1B Syringe Pump External control function

    01/06/2017

    External control function, including run/stop control input and direction control input can be enabled or disabled.1.In parameters setting interface, press and turn RES to highlight the external contr...

  • LSP02(10)-1B Syringe Pump Calibration

    01/06/2017

    The flowrate can be calibrated for more accurate dispense volume.1.In parameters setting interface, press and turn RES to highlight the Calibration line. Press RES to enter calibration interface.

  • LSP02(10)-1B Dual Channels Syringe Pump Pause time setting

    01/06/2017

    In infusion/withdrawal mode, the pause time is set in infusion parameters setting interface. In withdrawal mode, the pause time is set in withdrawal parameters setting interface. In continuous mode, t...

  • LSP02(10)-1B Dual Channels Syringe Pump Flowrate setting

    01/06/2017

    In parameters setting interface, turn and press RES to highlight the infusion flowrate or withdrawal flowrate setting line. Press RES to make the numerical value position highlighted. Enter the flowra...

  • LSP02(10)-1B Dual Channels Syringe Pump Volume setting

    01/06/2017

    In parameters setting interface, turn and press RES to highlight the infusion volume or withdrawal volume setting line. Press RES to make the numerical value position highlighted. Enter targeted volum...

  • LSP01(04)-1A Syringe Pump Calibration

    01/06/2017

    The flowrate can be calibrated for more accurate dispense volume.1.In parameters setting interface, press and turn RES to highlight the Calibration line. Press RES to enter calibration interface.

  • LSP01(04)-1A Syringe Pump Power up

    01/06/2017

    This option is only applicable when no dispense volume is selected. When power returns after an interruption the pump can resume operation (select Run) or remain stopped (select Stop).

  • LSP01(04)-1A Single Channel Syringe Pump Flowrate setting

    01/06/2017

    In parameters setting interface, turn and press RES to highlight the flowrate setting line. Press RES to make the numerical value position highlighted. Enter the flowrate required from the membrane ke...